Telonic TEST

Category: Knowledge Base RSA5000 Series

  • RIGOL´s new Solution for EMI pre-compliance tests

    RIGOL´s new Solution for EMI pre-compliance tests

    VNA Mode for Real-Time Spectrum Analysers

    Enhanced EMI measurement function and a simple user-friendly operating environment complete RIGOL’s real time spectrum analyser versions RSA3000N and RSA5000N.

    Electromagnetic compatibility [EMC] has become one of the most important topics in the last 23 years. EMC activities during design phase are absolutely mandatory to ensure the functionality of development and to protect the environment from electro smog. Each new development must be tested in a test laboratory in accordance with related standards in order to be CE certified. This certification is necessary to bring a new product to the market. Different standards are used depending what kind of product is designed. For example standard CISPR 22 (EN55022) needs to be passed for an “information technology equipment” [ITE] like a MP3 player or a modem (telecommunication). Each standard specifies frequency ranges for conducted and radiated emission of a product and defines maximum limits in dBµV for each analysis. Pre-compliance analysis is very important during design stage because it’s estimated that ~50% of products fail the first acceptance test. A second full compliance test is very expensive and time consuming. But also a re-design has a strong financial impact and uncalculated additional time is necessary which is a question of cost in the end as well. In order to avoid this scenario, initial EMC analysis and measurements are necessary at a very early stage of development and must be carried out throughout development, including pre-compliance checks. A survey1 shows that the average cost of EMC activities during development is at ~3-5% of whole development cost. However, if these activities will not be performed and the compliance test is failed, the cost of post- development can be arise up to 50% to 100% of the actual planned cost. For EMC analysis and pre-compliance control, RIGOL has been offering since years solution with spectrum analysers of the series DSA800 (including 6 dB filters and quasi-peak value detector [QP]) and PC Software S1210, as well as a near-field probe set NFP-3. This solution has now been extended and optimised in the RSA3000N/RSA5000N series. This makes it possible to simplify EMC analysis considerably and also offers even more advanced analysis capabilities.

    EMI Problematic

    EMC2 was implemented to generate a peaceful coexistence between transmitter and receiver of electromagnetic energy. Also unwanted transmitting and receiving are within this definition. Furthermore EMC should protect the recourse “electromagnetic spectrum”. Three main elements are defined in EMI (see figure 1).

    Source of Interference – Coupling Path – Susceptible Device

    The source of interference could come from e.g. parasitic back coupling effects from an amplifier or an electromagnetic transmission of an electric / electronic facility. Due to different coupling path the disturbance will be coupled to the susceptible device. This interference could cause different failure rates in functionality at the EMI receptor (susceptible device):

    1.) Performance degradation
    2.) Function failure
    3.) Equipment damage (worst case)

    To avoid such kind of scenario it is important to detect the root cause including coupling mechanism at DUT. The EMI interferences could be transmitted via different coupling path. For lower frequency ranges the interferences are mainly transmitted via line connection (like for power, data bus, analogue lines, RF connection) because the wavelength (λ)3 is bigger than the geometry and electrical and magnetically field are independent. Both fields don’t propagate and a quasi-static approach is necessary (analysis of inductive / capacitive elements). The line connections need to be tested if unwanted interferences occurring during / after activation of DUT at these lines. This kind of test is named conducted emission.

    When the wavelength of a DUT is equal or smaller than its frame, then magnetic and electric field are no longer independent and they start to combine to an electromagnetic field (EM). If e.g. the geometry has 6 * λ then the EM field is no longer only at the source, propagation happens and the DUT start to send out an EM wave (its acting as an antenna). Unwanted interferences with EM waves are named radiated emission. In the near field of the EM source a quasi-static approach is necessary.

    The occurrence of interferences is in most cases unforeseen and difficult to detect because different coupling effects could be root cause of interferences. For a device under test there are two main influences of interferences. First one is an interference which is related within one DUT and has an influence of functionality within the device under design. These interferences are intra-system disturbances which will be occurred e.g. crosstalk effects.

    On the other side, the DUT might be affecting a complete different independent system with EMI interferences which are inter-system disturbances. For example, at a radio it could be happened that a specific noise is hearable over the speaker when a mobile phone is ringing.

    Interferences could be coupled by:
    – Galvanic coupling (G-C): when two electric circuits using same impedance (e.g. same GND)
    – Inductive coupling (I-C): when two (or more) conductor loops with current flow influence each other with magnetic field
    – Capacitive coupling (C-C): when two electric circuits have different voltage potential – Electromagnetic coupling (EM-C): Wavelength is smaller than the geometry and EM wave radiated to environment of DUT and / or other sections of DUT.

    To find a root cause coupling path in a DUT is very difficult, because coupling path could also be a combination of different couplings (e.g. G-C and C-C. at the same time). To find a solution for a detected coupling path is comparatively simple.

    The next part of this document describes the test tool RSA5000N with the new EMI mode which is the tool for pre-compliance analysis.

    EMI Mode of Real Time Spectrum Analyser RSA5000N / RSA3000N
    The RSA5000N / RSA3000N are multifunctional spectrum analyser including a general purpose spectrum analyser [GPSA], a real time analyser [RTSA], a vector signal analyser [VSA], vector network analyser [VNA] and an EMI mode for pre-compliance tests. Therefore these devices are the best combination of high performance and flexibility and an optimal test solution for EMI pre-compliance analysis. For some EMI measurements, the real time mode could be a big benefit for some analysis to get a better overall picture of the problem or see more behaviour of DUT. The GPSA mode could be used to measure single components of a DUT like an amplifier and with VSA mode, the modulation quality or BER of an RF output could be measured e.g. during susceptibility stress.
    The new EMI mode offers a lot of new functionalities and advantages for pre-compliance tests and will be described in detail:
    Scan Table
    During the conformity check (conducted and radiated emission / susceptibility), a test receiver [TR] is generally used in a test laboratory. However, a TR has several disadvantages. It is a very expensive test tool and only usable in the field of EMC. A spectrum analyser [SA] on the other hand, measures faster and can be used for different applications and is significantly cheaper, especially at RIGOL. However, a TR has significant advantages in EMC measurements comparing to an SA. TR using a pre-selection which allows to use a different dynamic range for each partial measurement. In addition, the desired frequency resolution (usually RBW/2 and sometimes RBW/4) is easily achieved in a TR. The SA fulfils the most of CISPR 16.1 requirements and the results are a very good approximation, which is absolutely sufficient to perform pre-compliance tests. A TR fulfills CISPR 16.1 for 100% and can also be used for compliance tests. RIGOL´s new EMI solution in the RSA5000N / RSA3000N compensates most of the disadvantages with the scan table. The most used frequency ranges inclusive the correct 6 dB RBW filter for EMC tests are pre-defined in 10 individual ranges (see figure 2). All pre-defined settings can be modified / stored according to the own need. In order to compensate these disadvantages in a SA, the EMI solution of RIGOL stores the basic setting of the most important frequency bands, including the 6 dB RBW required, in a scan table in advance (see figure 2) in individual areas [Ranges]. All parameters can be adjusted separately per area according to your own wishes.

    For example, it is possible to use up to 10,000 measurement points per area. The basic settings of measurement points refer to a frequency resolution of RBW/2. For a higher resolution, different ranges can be combined for one test. With a combination of e.g. 3 ranges it is possible to use 30.000 measurement points over the whole frequency to reach expected resolution if necessary. The integrated pre-amplifier (option) and internal attenuator (0 to 50 dB) can be varied per range as well. With combination of different ranges it is possible to increase the dynamic range of the complete measurement without interrupting the measurement itself, because the dynamic range can be adjusted for each single range. The disadvantage of frequency resolution and dynamic range in an SA can be compensated in EMI mode. Additionally to the standard 3 dB RBW filter, the EMI solution contents also the integrated 6 dB filters (200 Hz, 9 kHz, 120 kHz and 1 MHz). Especially RSA5000N providing real time and vector signal analysis capabilities, which is not the case with a TR.
    Integration of Limits According to Standards
    After the desired frequency range has been selected, you can load the pre-defined and stored limit lines of the desired standard (e.g. EN55022, Class B, AV or/and QP) and display it on the fully logarithmic trace display. A measuring curve can be assigned to each limit. It is possible to activate a measurement curve with a separate detector setting for each pass/fail limit. Different measuring curves can also be measured at the same time. For each limit line it is possible to activate an additional security margin, which is considered in the Pass/Fail viewing. Especially in the pre-compliance check the result should be at least 5-6 dB below the defined limit line to ensure, that the compliance check will be passed.
    Advanced Meter/Additional Detectors/Corrections
    Another advanced measurement mode is the measurement meter. It is possible to activate up to 3 meters in parallel. Each of them can be used with different detectors and separate limit values. For example, the peak value detector [peak] can be set for first meter, second meter is set to QP detector and third meter to the new CISPR Average [C-AV] detector. The respective measuring meters can be set to the peak under focus via meter marker. The meters measures continuously. This means, for example, with an elevated peak value, the DUT can be reworked and the influence on this value can be immediately represented. A design improvement is immediately visible. With using three meters in parallel it is possible to see the peak measurement (worst case) but also the repetition pulse rate of this signal over a certain time with QP detector can be displayed. In addition a weighted average detector can be used for pulsed sinus-shaped signals with low repetition rate (C-AV4) which displays the maximum linear average value during the measurement.
    The meters can be coupled with a signal which is selected in the signal table. An additional fine-tuning of the meter frequency is not necessary. So each peak in signal table can be selected and measured by marker, as well. Alternatively, the meters could be coupled to a marker. For conducted emission tests additional components are used (e.g. line stabilization network (LISN) transient limiters, external attenuator, etc.). Each of the components has a direct influence of test result over frequency range. Therefore, it is possible to activate different corrections in RSA3000N / RSA5000N EMI mode. During the test, the correction will be respected in the result. Correction values can be stored internal the instrument, as a * .csv file. These can also be generated in the PC and loaded later on into the analyser.

    Measurement
    Different conditions can be set for the measurement. On the one hand, variable traces can be recorded at the same time with one or more detectors. Alternatively, only the worst case situation (peak detector) can be recorded. Afterwards each peak in the signal table will be measured with QP and C-AV detector, as well. It is no longer necessary to perform a whole trace with QP and C-AV (duration can be several hours) because these detectors are only interesting when peaks are visible and the test engineer has the information in some minutes and saves a lot of time during pre-compliance evaluation.

    The number of readings in the signal table can be pre-defined. In order to get a detailed view of a peak value, a zoom is available to enlarge a detailed view of a peak. After analysis, the view can be zoomed out to get back the original view.
    In case that the full screen of the trace is required, the measurement meter and the signal table can be turned off on the screen.

    Documentation
    In the EMI mode of RSA5000N / RSA3000N series, the measurements can be stored in different ways. For example, the complete signal trace can be stored as a * .csv file or a test report including settings, limits , an image of the graph including meter result and the signal table can be created as HTML or *.pdf in the analyser itself or on an external USB stick. It is possible to modify the header of the test report with information like temperature / humidity, name of operator or test location, etc.
    The next part of this document will go through some measurement examples with conducted / radiated emissions and near field probing.
    Conducted Emission
    As already mentioned, conducted emission is mainly measured up to 30 MHz at lines (e.g. power, analogue / digital signals, etc.). For conducted emission test on power line it is necessary to supply the DUT with power. At the same time the disturbances from DUT shall be measured. The problematic is, that also from the power supply disturbances could be send out which shall not be measured. Another aspect is the impedance matching to avoid a wrong result based on an impedance mismatch. The solution for this problematic is the usage of a line impedance stabilization network [LISN]. A LISN blocks all interferences (via inductive component) from power supply to SA, but supplies a DUT with power. Depending on inductive size, a LISN can be used for DC (5 µH) or AC (50 µH) power source. The power will be blocked on the way to SA (via a capacitance) but the DUT disturbances won’t be blocked and transferred to the SA. The impedance at SA connection is 50 Ω.

     

    Two security mechanisms are important using a LISN. When a LISN or a DUT will be activated, then the electric charge of capacitance changing and current flows for a short period. This transient current can destroy the first mixer diode in a spectrum analyser. It is recommended to use a transient limiter between LISN and SA, to protect the analyzer5. For the first test it is also recommended to use an external attenuator for the test to see if high disturbances will be occur from DUT. The attenuator can be removed after first test, when the user be sure, that the disturbance peaks are at a normal level.
    In each standard it is defined how the test setup has to be performed. In figure 6 there is the physical connection of a test setup displayed:

    The ground connection of LISN (and DUT if possible) needs to be connected to the horizontal ground plane.
    In the example below the conducted emission test was performed on a power supply (Phase) of a test device using a 50µH AC LISN6. First test was performed with integrated LISN limiter and the second test in figure 7 was performed without limiter. The limit lines and detectors are selected for standard EN55022, Class B. The test result is including correction values of LISN. This test uses two limit lines. First one is for average detector and second (blue) limit line is for QP detector. For information only, the peak detector was also activated for final measurement, because the peak detector displays always the worst case value. This power supply is very close below the limit margin, but it passed. Higher peaks can be analysed via measure meter.

    Additionally to EMI analysis the frequency band can be analysed in real time [RT] as well. This mode uses a very fast FFT algorithm to calculate the frequency spectrum. The calculation time is smaller than one FFT frame. Therefore a seamless capture of time signal is possible and no signal information7 will be lost. In RT mode different displays can be used. For example within density mode it is possible, to see a signal repetition over test time in different colour grades and therefore it’s possible to see more details. Density or Normal view could be combined with spectrogram (time waterfall over frequency range) where different amplitude values are displayed in diverse colour grades (see figure 8).

    An additional strong tool is the combination of frequency view and power vs time (with bandwidth up to 40 MHz8) in parallel. For a SA it is only possible to view the frequency spectrum or zero span with the max bandwidth of used RBW but not both in parallel. In figure 9 the same DUT was tested with smaller frequency range and with normal spectrum in combination of spectrogram. For this measurement it’s possible to speed up test time to 100 µsec. The interference pulses are now visible in spectrogram. With a Z marker it is possible to measure the difference in frequency, amplitude and time of an interference signal.

    Near Field Probing
    Near field analysis is a very important part in pre-compliance analysis and root cause debugging. RIGOL offers with the set of near fields (NFP-3) 4 different H probes for failure detection (see figure 10). It is a special form of emission analysis, because the measurement is done in reactive nearfield range (up to 0.159 * λ)9. In near field magnetic and / or electric field needs to be measured separately because both have a phase shift of up to 90° and nearfield stores blind energy without radiation. For most radiated emission root causes can be detected with locating the near field strength.

    Tests on housing leakage could be detected as well as strong disturbance components like ribbon cables or LCD panels. Bigger probes are usable to find the area of interferences. Smaller probes can be used to detect the exact source.
    One approach for near field analysis is to use the spectrum analyser with linear frequency scale / no limit lines and test e.g. a frequency range of 30 MHz to 500 MHz with a RBW of 100 kHz and activation of pre-amplifier (internal attenuator: 0 dB) to get a first view the interference peaks of DUT (see figure 11).

    Radiated Emission
    According to the description at the beginning, propagation starts when λ is small to the geometry and E & H are no longer independent. Both are in-phase and propagate to the environment. Radiated emission is not easy to test. In a test laboratory a big hermetic test chamber is used where far field conditions are fulfilled and other influences (due to external components like a mobile phone, etc.) are not visible in the result. Far field condition starts with a distance of 4 * λ and the characteristic wave impedance is 120 * π Ω (= 377 Ω). The hermetic room has in most cases at the wall broadband absorber for impedance matching of EM wave to avoid reflections. The altitude of broad band antenna can be changed during a test to find the highest peak at DUT during the test10.

    In most cases such kind of test chamber is not available for pre-compliance analysis. Therefore alternative ways has to be evaluated.
    First of all, it is possible perform a free field measurement with the distance of radiated near field which is defined for 0.159 * λ to 4 * λ. In this region a far field evaluation is possible (with using correction values). To detect the test area attenuation11 from environment it is possible to perform a test with a counter antenna instead of DUT and the results will be compared with defined formula of theoretical value. The difference is the test area attenuation. Depending on the distance between antennas, variable correction factors are used. This test needs a very good conductive ellipsis area12. There shall be no conductive obstacles greater than 5 cm within the ellipsis (see figure 12). DUT and test antenna are located within the focus of ellipsis. It is necessary to perform an altitude variation of test antenna to detect maximum.

    This test is very time intensive and expensive. A more simple way of radiated emission analysis for pre-compliance test is, to use a TEM cell.

    A characteristic for radiation of antennas is the far field. According to the description above, the characteristic wave impedance ZF = E / H which is in far field, is 377 Ω. The idea of a TEM13 cell is the medium of a coax cable which is suitable for a TEM wave propagation. The characteristic wave impedance in a coax cable is independent of the mechanical structure and always 377 Ω. A TEM cell is displayed in figure 1314. A common use for pre-compliance analysis is, to use an open TEM cell with disadvantage to see influence from environment. But this could be evaluated as a reference measurement with deactivated DUT. The two bends on the roof are necessary for the matching, but results into a frequency limit (max frequency range) is the cut off frequency based on the generation of higher wave guide modes on these bends which have an influence on test results. This cell can be used for susceptibility stress of a DUT but can also be used for repeatable pre-compliance radiated emission approximations which are sufficient for pre-compliance analysis. The connection to a spectrum analyser needs a DC blocker because the lowest cut off frequency of a coaxial cable is 0 Hz (= DC) which could destroy a spectrum analyser. For impedance matching of septum15 it is necessary to use a 50 Ω Termination on second port to avoid reflections. The DUT is placed between septum and lower shielding plate. A test setup is visible in figure 14. The advantage of open TEM cells, failures can also be analysed in parallel with a near field probe to detect the root cause area of DUT.

    Susceptibility Analysis
    Near field probes, current probes and TEM cells can also be used for conducted / radiated susceptibility stress of a DUT. RIGOL’s RF generator series DSG800 or DSG3000B are an optimal tool in combination with those components, to stress a DUT with an amplitude modulation [AM] e.g. according to IEC 61000-4-316, A functionality during stress might be controlled via the oscilloscope MSO8000 to see e.g. influence to jitter of clock signals or wrong decoding of bus systems or to see crosstalk effects on strip lines. Conducted susceptibility could be done with a current probe over the related stress cable. Also the real time spectrum analyser can be used to measure, in parallel, sporadic interference reaction at DUT due to susceptibility stress. The real time functionality offers different trigger types. One is the frequency mask trigger [FMT] which is a defined field (different colour) and the RTSA measures only, when the conditions of the field is fulfilled (e.g. measure, when a signal is inside the FMT field, see figure 15).

    The quality of digital modulation of RF interface during absorption stress can be tested via RSA5000N VSA (see figure 16). Here it is possible to display also I and Q stream separately to see which part of transmitter is affected by the interference. Additionally a BER test could be performed during stress situation.

    Additionally, with the vector network analyser function [VNA] it is possible to measure out related antennas or other components to see if a possible degradation of performance is visible (see figure 17).

    Summary
    RIGOL offers a new dimension of pre-compliance testing with the new EMC solution for the series RSA3000N/RSA5000N. With the additional meter measurement method high peak values can be determined and improved at a very early design stage. This test solution eliminates the need for PC test software. The EMV analysis is also greatly simplified by the pleasant operation of the device (touchscreen or operator panel as well as USB mouse or USB keyboard). In addition, the real-time capacities (real-time is integrated as standard) of the RSA series provide an expanded insight into EMC analysis.

  • Why am I seeing the message “Intermediate Frequency Out of Range”?

    Why am I seeing the message “Intermediate Frequency Out of Range”?

    There is too much power being delivered to the instrument.

    The Spectrum Analyser is a Swept Superheterodyne design and does not have preselection filters. Therefore, the mixer is seeing all of the power (full 9kHz-1.5GHz frequency range) regardless of what you have set for the start and stop values.

    A fixed attenuator should be added  to the input before running any more tests. Perhaps 10dB to start. Then, scan quickly (100kHz RBW) from 9kHz-1.5GHz and look for peaks near 10dB or greater. These spikes are overpowering the mixer and could damage the instrument.

  • Setting up the RSA3000 and RSA5000 series spectrum analysers to capture the lowest level signals

    Setting up the RSA3000 and RSA5000 series spectrum analysers to capture the lowest level signals

    The lowest level signal you will be able to see captured on a spectrum analyser will be limited by the DANL or displayed average noise level so the lower the DANL the  better.

    To obtain the lowest DANL the RSA3000/RSA5000 series analyzers will need to be operating within these parameters and settings:

    Attenuation = 0 dB
    Sample Detector
    Trace Average greater or equal to 50
    Tracking  Generator set to off
    Normalized to 1Hz
    Operating between 20°C to 30°C
    Input Impedance set to 50  Ohms

  • My spectrum analyser displays 1st LO Unlock. What could be wrong?

    My spectrum analyser displays 1st LO Unlock. What could be wrong?

    1st LO unlock indicates a hardware problem. The instrument will need to be repaired. 

    Contact Telonic for service on this instrument.

  • Precompliance: Susceptibility testing

    Precompliance: Susceptibility testing

    To learn more about using RIGOL Instruments in EMC Precompliance testing go to our application site at:

    www.rigolemc.com

  • EMC Precompliance: Near Field Probes

    EMC Precompliance: Near Field Probes

    To learn more about using RIGOL Instruments in EMC Precompliance testing go to our application site at:

    www.rigolemc.com

  • Common HAM and Broadcast uses for a spectrum analyser

    Common HAM and Broadcast uses for a spectrum analyser

    1. Measuring the performance of a crystal filter
    Hams use crystal filters in their radios to enhance selectivity, but short of installing the
    filter there is really no easy way to test one. A spectrum analyser with a tracking generator
    makes it fairly easy. Simply feed the tracking generator into the filter input and look at the
    resulting response from the filter output.
    A nice additional feature on some spectrum analysers is the “N dB” feature that makes it
    easy to find the point where the response is N dB down from that at the filter center
    frequency.

    2. Testing and adjusting duplexers and diplexers
    Duplexers are used in repeater installations to allow a single antenna to be used for
    simultaneous receive and transmit on slightly separated frequencies. In order to do this the
    duplexer has to be adjusted so that it provides the required attenuation of the transmitter
    frequency so that it does not interfere with the receiver.
    Any repeater owner has to deal with tuning a duplexer at some point. A spectrum analyser
    with a tracking generator makes this easy.
    A diplexer allows the use of a single antenna on two different bands, and is a lot smaller
    than a duplexer. Lots of HAMs use diplexers in their cars to allow a single antenna to be
    used on both the VHF and UHF bands.

    3. Coaxial stub tuning and coaxial cable measurements
    Attach a piece of coaxial cable to the input/output of the tracking generator using
    a T connector. First, this shows you the frequency where the cable is 1/4 
    wavelength long. ¼ wavelength coaxial stubs are often used to reduce interference from
    nearby transmitters. You can trim a piece of coaxial cable until you reach the frequency
    you want just by cutting a small amount from the cable and watching the response on the
    screen.
    Another useful feature using the T-connector technique is that when you have determined
    the frequency where the cable is 1/4 wavelength long you can easily determine the
    physical length of the cable without actually having to measure it. All you need to know is
    the frequency where the cable is 1/4 wavelength long (read from the screen on the
    spec an) and the velocity factor of the cable which is a specification that can be looked up
    for any type of cable.
    Then the length of the cable (in feet) is calculated by the formula L = (246*Vf) / F, where
    F is the frequency in MHz.

    4. Antenna performance (return loss)
    The return loss of an antenna is a measure of how close its impedance is to that of the
    system driving it. You can measure return loss with the spectrum analyser and the tracking
    generator if you have a directional coupler. By attaching the reflected port to the spectrum
    analyser you can measure the return loss of an attached antenna, and you can calculate
    VSWR from return loss by the formula VSWR = (1+(10^RL/20)) / ((10^RL/20)-1) where
    RL is the measured return loss.

    5. FM and AM Radio Channel performance.. you can demodulate, see the frequency and
    amplitude deviation, etc..

  • Do I need to have the Advanced Measurement Kit to use UltraSpectrum?

    Do I need to have the Advanced Measurement Kit to use UltraSpectrum?

    There are available advanced measurements within UltraSpectrum. These measurements include ACP, TOI, and others. To retrieve these measurements, the instrument must have the advanced measurement kit (AMK part number) installed.
  • EMC PreCompliance Measurements

    EMC PreCompliance Measurements

    RIGOL Spectrum Analysers are utilized in many applications, but EMC PreCompliance is of particular interest to our customers. All our applications content related to EMC can be found at:

    www.rigolemc.com

  • Using a Tracking Generator

    Using a Tracking Generator

    A great blog for reference on Spectrum Analyser work. This entry from Ken Wyatt talks about using a tracking generator. Find the content here.
  • Is a Preamp necessary with EMI work?

    Is a Preamp necessary with EMI work?

    It’s generally recommended to use a preamp with a spectrum analyser when performing EMI and EMC pre compliance tests. The preamp is used to amplify the acquired signal above the noise floor for low level signals. So it’s very useful for EMI work.

    Final determination will come down to what energy levels will be tested.

  • What operating system is used on the RSA3000/RSA5000 series spectrum analysers?

    What operating system is used on the RSA3000/RSA5000 series spectrum analysers?

    The operating system used for the RSA3000/RSA5000 series spectrum analysers is Linux v3.12.0, 32 bit.