  • Detector use cases

    Detector use cases

    This document highlights some of the common uses for Spectrum Analyser Detector Types.

  • S1210 Changing the Workspace path

    S1210 Changing the Workspace path

    When the S1210 application is started up a default Workspace path and demo file is created. In the example below the demo Workspace file has been created in C:/Users/ruby/AppData/Local:



    The S1210 application now shows the associated demo tests that be selected and you can connect to the spectrum analyser by clicking on Device:



    A different Workspace path can be selected if desired. In the example below the path has been set to C:/Temp/S1210: 



    No tests have been created so there are now no tests listed in the example below. And, without a test you will not be able to connect with the spectrum analyser when you click on Device: 



    To create a new test select File and then New. In the example below test_0 has been created. Now that a test exists you can select Device to connect to  the spectrum analyser. 

  • How to Load an RSA3000 License Option

    How to Load an RSA3000 License Option

    Loading a License on an RSA Spectrum Analyzer is simple.

    First, you receive the PDF file with the License key to install:


    Enter the license Key and the Instrument serial number into the website

    https://licenseen.rigol.com/CustomerService/ProductRight_EN


    The verification code is the text to the right.

    For the RSA instruments this generates a .LIC file that you can load onto a USB stick.

    The files always have the filename equal to the Serial number, so only load one at a time to the USB stick to avoid changing the filenames or replacing the files.

    Once you move the LIC file to a USB stick you can start the RSA Spectrum Analyzer and then insert the USB drive into any of the USB ports.

    On the instrument hit the FILE button then use the touchscreen or mouse to navigate to the USB stick and the LIC file that you wish to load. Highlight the file so that it is in Blue and then go to the second page on the FILE menu using the down arrow. The screen should now look like this:



    Once there, push the IMPORT LICENSE menu item.

    This will complete the process.

    At any time you can verify what is installed in an RSA instrument by pushing SYSTEM, ABOUT SYSTEM, and then selecting OPTION INFO. This popup appears:



    <

    Rigol USA Web Site

     

    For Assistance in North America:

    Call Us @ 877.474.4651

    Email Us @ help@rigol.com

    www.rigolna.com

    For Assistance in Europe:

    Call Us @ +49 (0)89-8941895-01

    www.rigol.eu

    For Assistance in China:

    No.8 Ke Ling Road,

    Hi-tech Development Zone,3

    Suzhou,China 21516

    info-intl@rigol.com

    Sales: +86-512-69585066

    Fax: +86-10-80720067

    www.rigol.com

    For Assistance in Other Countries:

    info-intl@rigol.com

    int.rigol.com


    Warranty Terms
    Activate

  • EMI Test System Software Getting Started Guide

    EMI Test System Software Getting Started Guide

    The Rigol EMI Test System Software is designed to make precompliance scanning with a Rigol DSA family of Spectrum Analyzer easy. 

    This Note is an addendum to the Software User’s Guide. We will set up a prescan, create an amplitude correction table, set limit lines, and perform a final scan. 

    Initial Setup: 

    –    Download the latest version of UltraSigma and the EMI Test System Software from the Rigol website. 
    –    Connect your instrument (USB, LAN, or other) and set up the physical test environment (cabling, Device-Under-Test, etc..) 
    –    Start UltraSigma 
    –    USB connected instruments should appear in the UltraSigma Resource tree as shown below: 


    –   Right-click on the resource and select EMI Test System 
      

    –    The EMI Test System will perform a screen grab of the Spectrum Analyzer and open: 
     
          

    Configure Prescan Parameters 
     
    –    Press Edit Scan to open the scan parameters window 
     
     
    –    A Pre Scan consists of subrange elements. You can use the Pre Scan Configure window to set the overall Scan Start, Scan Stop, Range, and Detector type to use throughout the test: 
     
          
    Pre Scan Configuration sets the overall test parameters used in each subrange scan. Using a Positive Peak detector will allow you to perform a faster scan and identify problem areas: 
     
    In this example, we have configured the first subrange to sweep from 9kHz to 10MHz  at an RBW of 1MHz, no attenuation, and no preamplifier. 
     
      
     
    –    Subrange Configuration sets the parameters used for each subrange step: 
     
    In this example, we add a second step that goes from 10MHz to 30MHz in  at an RBW of 300kHz: 
     
      
    Build successive subrange configurations using Range Start and Range Stop values and use the Add, Insert Before, After, and Delete to arrange the scan 
     
    –    Press OK after you have built the Pre Scan list: 
     
      
     
    –    Press Prescan to collect prescan data: 
     
      
          
    After the scan is complete, you should have a graphical display of each subrange stitched together onto one graph: 

      
     
    –    Now you can add amplitude correction by pressing Ampt Corr: 
     
        
    –    You can select Antenna, Cable, Other, or the User tab to specify which correction table to edit.  
     
    –    Enter the frequency point and corresponding  correction level. Press Add to add the correction point to the table. Repeat until the full correction table is built. 
     
    –    To apply a correction table, select the type (Antenna, Cable, Other, or User) from the check boxes and press Correction: 
      
     
    –    You can save the correction files for later use by pressing the Save button. 
     
    You can add limit lines by pressing Peak Search to bring up the Peak Search window: 

          
    –    You can add points to the limit line one-at-a-time by entering the frequency and level, then press add. Here, we have built a limit line comprise of four sections. You can also save the limit lines to recall them later by pressing Save: 
     
      
    –    Use History to recall saved limit lines 
     
    –    Press OK to apply the limit lines 
     
     
    –    All of the peaks that have failed the limit test are shown at the bottom of the display: 
     
          
    –    You can find specific peaks by clicking on the Peak number: 
     
      
    –    You can also change the Vertical and Horizontal Scales by selecting the scales from the Y Unit and X Axis drop down boxes: 
      
     
    –    You can print out a report of scanned data by pressing the Report button and filling in any information that you need. 

     
    –    If you have found areas that fail the limit lines, you can save time by performing another prescan. This time, set the start and stop frequencies to encompass the failing peak(s). 
     
    –    In this more focused scan, you may also want to use a Quasi Peak detector and FCC RBW settings (if your analyser is equipped with those options.

  • RSA3000, RSA3000E and RSA5000 Alternative Factory Reset and Firmware Upgrade

    RSA3000, RSA3000E and RSA5000 Alternative Factory Reset and Firmware Upgrade

    To reset the RSA3000, RSA3000E and the RSA5000 series to factory settings without using the System menu:

    1) Power cycle the instrument
    2) During the boot sequence, quickly and repeatedly press the Back button directly below the keypad on the instrument. 


    To upgrade the firmware on the RSA3000, RSA3000E and the RSA5000 series to factory settings without using the System menu:

    1) Power cycle the instrument
    2) Insert a flash drive into the front of the instrument with the latest firmware version loaded onto the root directory of the drive. 
    3) During the boot sequence, quickly and repeatedly press the Preset button in the upper right hand corner of the instrument.

  • RSA3000/3000E/5000 series Open Source Acknowledgment

    RSA3000/3000E/5000 series Open Source Acknowledgment

    RIGOL’s RSA3000/3000E/5000 series product uses a number of open source software packages. This document lists the open source software packages used in RSA3000/3000E/5000 series product. 

    https://beyondmeasure.rigoltech.com/acton/attachment/1579/f-be0541d6-bc36-481f-b448-ec059be1b26e/1/-/-/-/-/RSA3000_3000E_5000.pdf

  • How to save a screen shot on the RSA3000 and RSA5000 series at the push of a button.

    How to save a screen shot on the RSA3000 and RSA5000 series at the push of a button.

    In order to take a screen shot on the RSA5000, RSA3000 or RSA3000E and have it automatically save to your flash drive follow the steps below.
    1) Insert flash drive into the instrument and confirm that you can identify it in the File Explore menu.
    2) Press the Save button
    3) Press Quicksave Path
    4) Press Set Path
    5) Select your flash drive
    6) Press OK.
    This will change the quick save buttons path from saving a screen shot internally to your flash drive.

  • S1210 – Using dBuV units successfully

    S1210 – Using dBuV units successfully

    Recent versions of DSA815 Firmware and S1210 software improved the operation of the amplitude units in the software. These versions are:

    Instrument Firmware – 00.01.19

    1210 Software – 05.00.07

    On occasion it is still possible to show a data offset due to changing units such as this:

    This is usually caused by using the instrument from the front panel and setting the amplitude to dBuV manually. The software is expecting the instrument to be in default dBm before it runs. The best way to for data to be correctly processed is to hit the green PRESET button if you have been changing settings on the instrument between software runs.

    With the factory preset active the data is correct in dBuV:

  • S1210 Final Scan

    S1210 Final Scan

    Once a pre-scan has been done there are 2 ways to add a marker:

    Double click on the graph in the Measure window. That will create a marker in the Marker Table:


    Or right click in the marker table area to add a marker:


    When the markers are set correctly you can right-click on the marker to select Add to Peak Table:


    Now when you tab to the Peak Table you see the added Peak (in addition to the Peak Search peaks that have been located if you clicked on Peak Search).

    Now, in the Test menu select Final Scan Start. The final scan will update the Final Svan Levels for all the items in the Peak table using the settings on the scan config:

  • Saving and Using Test Setups in EMI mode

    Saving and Using Test Setups in EMI mode

    Relative measurements are critical in EMI: How does the good board compare to the failed board? How does this look with the CISPR Average detector or against another limit line? Is that coming from the device or the environment? Learn how to create, organise, and open complete test setups for comparison and data recording. Setups can include data traces and instrument configuration details.

  • EMI Automated Test Setup

    EMI Automated Test Setup

    Make EMI measurements simpler and faster by automating the process. Combine scans, searches, and interactive measurements. Also conduct further analysis on signals of interest with a single button on the same instrument engineers use for advanced debugging of emissions and signals.

  • How to use Limit Lines in EMI mode

    How to use Limit Lines in EMI mode

    Access the built in limit lines or configure your own for custom testing. Use the limits and margins when searching for signals of interest or for interactive measurements using multiple detectors.