Telonic TEST

Author: James

  • Programming Example: SDS Oscilloscope screen image capture using Python

    Programming Example: SDS Oscilloscope screen image capture using Python

    Here is a brief code example written in Python 3.4 that uses a socket to pull a display image (screenshot) from a SIGLENT SDS1000X-E scope via LAN and save it to the local drive of the controlling computer.

    NOTE: This program saves the picture/display image file in the same directory that the .py file is being run from. It will overwrite any existing file that has the same name.

    Download Python 3.4, connect a scope to the LAN using an Ethernet cable, get the scope IP address, and run the attached .PY program to save a bitmap (BMP) image of the oscilloscope display.

    You can download the .PY file here: Python_Socket_SDS_SCDP.zip

    Tested with:

    Python 3.4
    SDS1202X-E
    SDS1104/1204X-E
    SDS2000X-E Models
    SDS5000X Models

  • X-Y Display mode on Scopes – additional features with RIGOL DS1000Z-series oscilloscope (DS1074, DS1074Z-S, DS1104Z, DS1104Z-S)

    X-Y Display mode on Scopes – additional features with RIGOL DS1000Z-series oscilloscope (DS1074, DS1074Z-S, DS1104Z, DS1104Z-S)

    New DS1000Z series offers more in X-Y mode:

    In this mode, the Rigol DS1074Z oscilloscope changes 2 of the 4 channels from voltage-time display mode to voltage-voltage display mode. The phase deviation between two signals with the same frequency can be easily measured via Lissajous method. The figure below shows the measurement schematic diagram of the phase deviation.

    To access the Lissajou mode of your oscilloscope press the horizontal menu button, and then change the time scale on the screen to X-Y mode. From here you can choose between CH1-CH2, CH1-CH3, CH1-CH4, CH2-CH3, CH2-CH4 and CH3-CH4 as your inputs.

    NOTE: RIGOL’s new DS1000Z-series offers two features in X-Y mode: you can still see a condensed normal signal display (Y-T) at the top margin of the screen and you can also access the Trigger Menu in X-Y mode.

  • Which oscilloscope probe is best for my application

    Which oscilloscope probe is best for my application

    Selecting the correct oscilloscope probe is very important for making quality measurements. If the probe performance is not adequate or if you did not select the proper probe for your application, you will see distorted or misleading signals on your oscilloscope. There are two kinds of probes, passive and active. Active probes requires power while passive probe do not. For general-purpose measurements(<600MHz), passive high-impedance resistor divider probes are good choices in general. For high-frequency applications less than 600 MHz, active probes are the way to go. They cost more than passive probe and their input voltage is limited, but because of their significantly lower capacitive loading, they give you more accurate insight into fast signals. Selecting the right probe is the first step for your application towards correct measurements. Below is the RIGOL scope probe table which can help you to select a proper probe for your application.

    Type
    Model
    Number
    Attenuation
    Ratio
    Bandwidth
    Input R
    Max. Input voltage
    RIGOL scope Compatility
    Recommended
    Applications
    Passive high resistance probe
    RP2200
    1:1 or 10:1

    1X:DC~7MHz
    10X:DC~150MHz

    1X:1MΩ±2%
    10X:10MΩ±2%
    1X:CAT||150V AC
    10X:CAT||300V AC
    DS1000 series,DS2000 series,DS4000 series,DS6000 series small signal test(1X),
    General purpose test
    RP3300A
    10:1
    10X:DC~350MHz 10X:10MΩ±2% 10X:CAT||300V AC
    DS1000 series,DS2000 series,DS4000 series,DS6000 series
    General purpose test
    RP3500A
    10:1
    DC~500MHz
    10MΩ±2% CAT||300V AC DS1000 series,DS2000 series,DS4000 series,DS6000 series General purpose test
    RP5600A
    10:1
    DC~600MHz
    10MΩ±2% CAT||300V AC DS4000 series,DS6000 series General purpose test
    Passive low resistance probe
    RP6150A
    10:1
    DC~1.5GHz
    500Ω±10Ω CAT|7V AC DS4000 series,DS6000 series High frequency signal ended small signal test
    Passive high voltage probe
    RP1300H
    100:1
    DC~300MHz
    100MΩ CAT|2000V(DC+AC)
    CAT||1500V(DC+AC)
    DS1000 series,DS2000 series,DS4000 series,DS6000 series High voltage test
    RP1050H
    1000:1
    DC~50MHz
    100MΩ DC:0~15KV DC
    AC:pulse≤30KVp-p
    AC:sine wave≤10KVrms
    DS1000 series,DS2000 series,DS4000 series,DS6000 series High voltage test
    Active differential probe
    RP7150
    10:1
    DC~1500MHz
    Differential mode:50kΩ±2%
    Single ended mode:24kΩ±2%
    ~30V(DC+AC) DS4000 series,DS6000 series Differential/Single ended high frequency signal test
    High Voltage differentia RP1025D X20,X50,X200 25MHz Differential :4MΩ/1.2pF
    Single ended :2MΩ/2.3pF
    Atten X20:
    ≤140Vpp,(45Vrms or DC)
    Atten X50:
    ≤350Vpp,(110Vrms or DC)
    Atten X200:
    ≤1400Vpp,(450Vrmsor DC)
    DS1000 series,DS2000 series,DS4000 series,DS6000 series High voltage differential/Single ended signal test
    RP1050D X100,X200,X1000 50MHz Differential:100MΩ/1.2pF
    Single ended mode>:50MΩ/2.3pF
    Atten 100:
    ≤700Vpp,(230Vrms or DC)
    Atten X200:
    ≤1400Vpp,(460Vrms or DC)
    Atten X500:
    ≤3500Vpp,(1140Vrmsor DC)
    Atten X1000:
    ≤7000Vpp,(2300Vrmsor DC)
    DS1000 series,DS2000 series,DS4000 series,DS6000 series High voltage differential/Single ended signal test
    RP1100D X100,X200,X1000 100MHz Differential :100MΩ/1.2pF
    Single ended mode:50MΩ/2.3pF
    Atten 100:
    ≤700Vpp,(230Vrms or DC)
    Atten X200:
    ≤1400Vpp,(460Vrms or DC)
    Atten X500:
    ≤3500Vpp,(1140Vrmsor DC)
    Atten X1000:
    ≤7000Vpp,(2300Vrmsor DC)
    DS1000 series,DS2000 series,DS4000 series,DS6000 series High voltage differential/Single ended signal test
    Model
    Number
    Gain Bandwidth Gain Accuracy Max.Input Current
    RIGOL scope Compatility
    Recommended
    Applications
    Current Probe RP1001C 0.01V/A,0.1V/A DC to 300KHz 100mV/A:±3% ±50mA
    (50mA to 10A peak range )
    10mV/A:±4% ±50mA
    (500mA to 40A peak range)
    100mV/A:±15% Max.
    (40A to 100A peak range)
    DC:±100A
    AC P-P:200A
    AC RMS:70A
    DS1000 series,DS2000 series,DS4000 series,DS6000 series Current test
    RP1002C 0.05V/A,0.5V/A DC to 1MHz 500mV/A:±3% ±20mA
    (20mA to 14A peak range )
    50mV/A:±4% ±200mA
    (200mA to 100A peak range)
    50mV/A:±15% Max.
    (100A to 140A peak range)
    DC:±100A
    AC P-P:200A
    AC RMS:70A
    DS1000 series,DS2000 series,DS4000 series,DS6000 series Current test
    RP1003C 0.1V/A DC to 50MHz ±1.0%rdg ±1mV, ≤30A
    ±2.0%rdg , 30A to 50A peak non-continuous
    AC P-P:50A peak,non-continuous
    AC RMS:30A
    DS1000 series,DS2000 series,DS4000 series,DS6000 series Current test,must order RP1000P power supply
    RP1004C 0.1V/A DC to 100MHz ±1.0%rdg ±1mV, ≤30A
    ±2.0%rdg , 30A to 50A peak, non-continuous
    AC P-P:50A peak,non-continuous
    AC RMS:30A
    DS1000 series,DS2000 series,DS4000 series,DS6000 series Current test,must order RP1000P power supply
    RP1005C 0.01V/A DC to 10MHz ±1.0%rdg ±1mV, ≤150A
    ±2.0%rdg , 150A to 300A peak
    AC P-P:300A peak,non-continuous
    500A Peak,pulse width ≤30us
    AC RMS:150A
    DS1000 series,DS2000 series,DS4000 series,DS6000 series Current test,must order RP1000P power supply
  • Using a Rigol product key to generate a Rigol software licence code.

    Using a Rigol product key to generate a Rigol software licence code.

    Please note…

    • You must enter the product key exactly as it appears on the certificate.
    • The serial number is made up of capital letters and numbers only – There are no lower case letters and no spaces.
    • Please take care when entering the serial number! If you enter an incorrect but valid serial number, the generated software licence code will not be accepted by your instrument.
    • You might find that entering the software licence code into your instrument via the front-panel is rather tedious. Please note that it is possible to enter the code remotely from a PC, using the SCPI control panel within Ultra Sigma. Copying the software licence code from the webpage generator to Ultra Sigma reduces the risk of error. For further information, please refer to the Ultra Sigma Help Document and the Programming Guide for your instrument. Both of these documents and the Ultra Sigma software download can be found on the webpage for your instrument, under the ‘DOCUMENTS, SOFTWARE & VIDEOS‘ tab.
    • For some models it is possible to download the generated software licence code as a ‘.lic’ file, save it to a USB memory stick, then import it into your Rigol instrument from the USB memory stick. For further information, please refer to the User Guide for your instrument.
  • RIGOL scope has a message on screen saying WAIT (in green). All channels ON but I see no signal being displayed. Is this effect to do with triggering?

    RIGOL scope has a message on screen saying WAIT (in green). All channels ON but I see no signal being displayed. Is this effect to do with triggering?

    Yes, you’re right: a WAIT status message is normal behaviour for the Trigger when the scope is not being given quite the right signal for its trigger to fire:
    – You’ll probably remember on all analogue scopes, there are at least two modes: Auto and Normal (sometimes also Single).
    – Normal is where the trace waits for each trigger before it starts to trace across the screen (we all remember the blank trace on old fashioned scopes, until you put it on Auto!).
    – Same with all digital scopes… and because all channels work off same trigger, you won’t see any channels at all.

    All RIGOLs indicate this Normal “untriggered” state by putting up the WAIT announcement. It just means the trigger condition is not being met.

    There are three things you can do here:
    1) Set your trigger mode to Auto
    2) Push the force button, which will invoke just one sweep manually
    3) If there is any readable signal, you can also keep it on Normal, and simply adjust the whatever is receiving signal, and then use the trigger threshold knob to ensure the dotted trigger threshold line on the screen is within the signal levels (just make sure the trigger is spy-ing on the right channel though!

    The advanced trigger types are best avoided too, as they might add more conditions, so just use Edge to begin with.)

    Glad to see you are having fun exploring your scope. Please let our RIGOL-UK team here at Telonic know if this seems right for your set up.

  • How can I use Ultra Sigma to grab data from my RIGOL Scope? E.g. CSV file for viewing in Excel?

    How can I use Ultra Sigma to grab data from my RIGOL Scope? E.g. CSV file for viewing in Excel?

    First of all, install and run Ultra Sigma,
    – Then make sure you can see your scope (Type shown in the Instruments window of Ultra Sigma)
    – AND that the scope response with its serial number to a *IDN? command. (Rt-click on instrument and choose SCPI Panel Control window, then click Send&Receive a *IDN?)

    Once this is OK, Change the setting shown as Base to Advanced (U’Sigma setting drop down next to the word Base – then change Base to Advanced)
    – Click Options – set Timeout to 12000 (not 2000) and Bytes To Read to 11024 (not 1024).
    – Enter in the command line the commands:
    i) :WAV:SOUR CHAN1
    ii) :WAV:MODE NORM
    ii) : WAV:DATA? and click Send&Read after each – a bunch of data should now appear in the window below
    – Select the second tab below, called Current Return Value (data window)
    – Right-click the data and choose Save Current Data To File-> Save for Byte -> Give your CSV file a name

    This CSV file can be opened directly in e.g. Excel, to graph data.

    NOTE: This grabs the SCREEN datapoints at the time of executing the :WAV:DATA? query. Works whether scope is running or not!
    To grab the MEMORY datapoints (which can be much longer and require different Options settings!!) you must first :STOP the scope and ensure your command is in :WAV:MODE RAW.

  • How to measure an RF Amplifier using a DSA-800 Series Spectrum Analyser

    How to measure an RF Amplifier using a DSA-800 Series Spectrum Analyser

    Solution: This document provides step-by-step instructions on using the Rigol DSA-800 series of Spectrum Analysers to measure the characteristics of an RF Amplifier.

    In addition to the DSA-800 Spectrum Analyser, you will need an RF Source, cabling, and adapters.

    Measure the amplifier

    1. Connect the RF generator output to the RF input of the instrument using the appropriate cabling and connectors.

    NOTE: If your instrument is equipped with a Preamplifier, you can enable it to lower the displayed noise floor by pressing the following sequence:

    Press AMP button > Down Arrow > RF Preamp On

    2. You can use the Auto button to center and zoom on the waveform. You can also use the Freq and Span buttons to manually manipulate the displayed data.

    Another option to center the waveform is to press Freq button > Peak → CF. This will automatically align the center of the display with the peak of the trace.

    3. Freeze the unamplified trace by pressing Trace > Trace Type > Freeze. You can use the Marker button to create a marker. This can be used to find the peak frequency and amplitude of the displayed waveform.

    4. Disable the RF Generator output.

    5. Disconnect RF generator from the instrument RF Input and connect it to the Amplifer input.

    6. Connect the Amplifier output to the instrument RF Input.

    7. Enable the RF generator.

    8. Enable a second trace to visualize the amplified signal by pressing the Trace button > Select Trace 2.
    9. Set the trace type to Clear/Write by pressing Trace > Type > Clear/Write. 10.You can use the Auto button or manually center the trace using the Freq,
    Span, and Amp buttons.

    11. Readjust the amplitude scale by pressing Amp > Auto

    12. You can enable an additional marker for the new trace by pressing Marker > Select the marker you would like to use

    13. Now, select the trace you want to mark by pressing Marker > Marker Trace
    > select trace of interest

    • Be sure Normal is selected to enable the marker

    14. You can also enable a marker table by pressing Marker > Down Arrow > Mkr Table ON. This allows a convenient way to compare markers and values between traces.

    15. Alternately, you can use the Trace Math function to create a Trace difference on the screen.

    16. Enable Trace Math by pressing Trace > Trace Math

    17. Set Function to A-B

    18. Set A = T1

    19. Set B = T2

    20. Set Operate > On

    21. Set Amplitude by pressing AMPL > Auto

    NOTE: New trace appears. This represents Trace 1 – Trace 2.

    22. Set Marker to Math Trace by pressing Marker > select Marker 3 23.Set Marker Trace to Math by pressing Marker > Marker Trace > Math

    23.Set Marker Trace to Math by pressing Marker > Marker Trace > Math

    • You can then move the marker to the smoothest portion of Trace 3

    • You can then move the marker to the smoothest portion of Trace 3.

  • How do I measure an RF Bandpass filter

    How do I measure an RF Bandpass filter

    This document provides step-by-step instructions on using the Rigol
    DSA-800 series of Spectrum Analysers to measure the characteristics of an RF
    Bandpass filter.
    NOTE: The DSA must have a Tracking Generator to effectively perform the
    following test.
    Normalize the trace (Optional)
    Many elements in an RF signal path can have nonlinear characteristics. In many
    cases, these nonlinear effects on your base measurements can be minimised by
    normalising the instrument.
    1. Connect tracking generator output to RF input using the same cabling that
    you will be using to test your device. Any element, like an adapter, used
    during normalization should also be used during device measurement as any
    changes to the RF signal path could effect the accuracy of the measurement.
    2. Enable the tracking generator by pressing the TG button > TG On
    3. Store the reference trace by pressing the TG button > Normalize > Stor Ref
    4. Enable normalization by pressing the TG button > Normalize > Normalize
    On
    Measure the filter
    1. Connect the tracking generator output to the filter input using the appropriate
    cabling and connectors.
    2. Connect the filter output to the instrument RF input.

    3. Set the tracking generator amplitude by pressing the TG button and the TG
    Amplitude. You can use the keypad or wheel to enter the correct value.

    NOTE: If your instrument is equipped with a Preamplifier,
    you can enable it to lower the displayed noise floor
    by pressing the following sequence:
    Amplitude button > Down Arrow > RF Preamp On
    4. Enable the Tracking generator by pressing the TG button > RF Source ON
    You can see the small bump in the figure below.

    Figure 1: Before Auto.
    5. You can use the Auto button to center and zoom on the waveform. You can
    also use the Freq and Span buttons to manually manipulate the displayed
    data.

    Figure 2: After Auto.
    6. You can now enable the Marker function to measure the bandwidth and
    attenuation or passband characteristics of the filter.
    7. Press Marker Fctn > N dB BW and set the function to the amplitude of
    interest. In this example, we are measuring the 3dB Bandwidth of our filter.

  • EMC Precompliance: Testing on a budget

    EMC Precompliance: Testing on a budget

    Solution: Almost any electronic design slated for commercial use is subject to EMC (Electromagnetic Compatibility) testing. Any company intending to sell  these products into a country must ensure that the product is tested versus specifications set forth by the regulatory body of that country. Here in the US, the FCC specifies rules on EMC testing. CISPR and IEC are also used throughout the world.

    To be sold legally, a sample of the electronic product must pass a series of tests. In many cases, companies can self-certify, but they must have detailed reports of the test conditions and data. Many companies choose to have these tests performed by accredited compliance company. This full compliance testing can be very expensive with many labs charging thousands of dollars for a single day of testing. Testing a product for full compliance can also require specialised testing environments and no changes can be made during the testing. Any failures in compliance testing require that the design heads back to Engineering for analysis and possible redesign. This can cause delays in product release and an obvious increase in design costs.

    One method to lower the additional costs associated with EMC compliance is to perform EMC testing throughout the design process well before sending the product off for full compliance testing. This pre-compliance testing can be very cost effective and can be tailored to closely match the conditions used for compliance testing. This will increase your confidence in passing compliance the first time through, lower your test costs, and speed your time to market.

    Measuring Radiated EMI
    The most simple form of pre-compliance measurements for radiated emissions can be performed using a spectrum analyser, like the Rigol DSA-815 (9kHz to 1.5GHz), and near field electric (E) and magnetic (H) probes.

    Figure 1 and 2: Near field E and H probes.

    The most simple test is to configure the DSA to use the peak detector and set the RBW and Span for the area of interest per the regulatory requirements for your device. Then select the proper E or H probe for your design and scan over the surface of the design.

    Probe orientation (rotation, distance) is also important to consider. The probes act as an antenna, picking up radiated emissions from seams, openings, traces, and other elements that could be emitting RF. A through scan of all of the circuit elements, connectors, knobs, openings in the case, and seams is crucial.

    Figure 3: Using a near field H probe to test a power supply.

    For the first pass, configure the spectrum analyser to use the peak detector. This will provide you with a “worst case” reading on the radiated RF and it is the quickest path to determining the problem areas. Larger probes will give you a faster scanning rate, albeit with less special resolution.

    Once you have a good idea of your problem areas, you can get more detail by implementing a few common techniques. If you can, select a spectrum analyser that has the standard configuration used in full compliance testing. This includes a Quasi-Peak detector mode, EMI filter, and Resolution Bandwidth (RBW) settings that match the full test requirements specified for your product.

    This type of setup will increase testing time but should be used on the problem areas. A full compliance test configuration will mean your pre-compliance testing will provide a greater degree of visibility into the EMI profile of your design.

    On many instruments, you can also store cable and antenna correction factors that will allow you to see the true signal, without the added errors from the setup.

    The next step in radiated testing includes using antennas in place of the near field probes, a rotating platform for the equipment under test (EUT), and can include a special room that minimises environmental factors (semi-anechoic). These setups are beyond the scope of this document, but there are references at the end that provide good references for the details of the setup.

    Figure 4: Compliance Setup for Radiated Emissions Testing

    Measuring Conducted EMI
    Conducted EMI testing requires analysing the RF energy that is coupled from the instrument or test circuit to the main power line it is connected to.

    Like Radiated EMI, Conducted EMI is also measured using a spectrum analyser, but it also requires a transient limiter and a Linear Impedance Stabilisation Network (LISN). A LISN isolates the power mains from the equipment under test, isolates any noise generated by the EUT,  and couples the signals generated by the EUT to the spectrum analyser.

    Figure 3: Standard Conducted Emissions pre-compliance setup via a transient limiter. Note: be sure to choose a limiter suited to your own system and testing regime and do your own checks first: transients can easily exceed the input rating of your analyser (e.g. at switch-on – it is good practice to avoid switching on the LISN with the analyser connected) A modern DSA’s internal level indicator is not intended to warn against fast transients and internal attenuators do not provide greater input circuit protection – if in doubt do your own measurements first to ensure you always operate within the stated input limit envelope.

    As with emissions testing, the best start is a scan over the frequency range of interest using the peak detector on the spectrum analyser. Then, performing a quasi-Peak scan using the EMI filter for the problem areas. This will minimise test time while maintaining a high degree of confidence in your test.

    Summary
    EMC Compliance testing is mandatory for the majority of electronic products that are slated for sale throughout the world. For the cost of 1 day of compliance testing, you can have a pre-compliance setup that you can use to continually monitor and improve your design. This will help speed product development and  save the company money.

  • AMK: What do you get with the DSA Advanced Measurement Kit ? (DSA1000-AMK)

    AMK: What do you get with the DSA Advanced Measurement Kit ? (DSA1000-AMK)

    Advanced Measurement Functions
    Option DSA1000-AMK provides plenty of advanced measurement functions including T-Power (Time domain Power), ACP (Adjacent Channel Power), Chan Pwr (Channel Power), OBW (Occupied Bandwidth), EBW (Emission Bandwidth), C/N Ratio, Harmo Dist (Harmonic Distortion), TOI (Third Order Intermodulation) and Pass/Fail. The measure mode can be Single or Continuous and you can control the measurement status: Restart, Pause or Resume.

    Pressing the front panel key Meas, the corresponding menu will appear on the right of the screen. Press Meas Fctn and choose a measurement function. The screen will be divided into two windows. The upper one is for basic measure, displaying sweep trace, and the lower one shows the measurement results.
    1. T-Power (Time domain Power)
    Enables the Zero Span Mode and calculates the power within time domain. The measurable power types are Peak, Average and RMS.

    2. ACP (Adjacent Channel Power)
    Measures the power of the main channel and the adjacent channels and calculates the power difference between the main and each of the adjacent channels. When enabled, both the span and resolution bandwidth of the analyser are adjusted to be smaller automatically.

    3. Chan Pwr (Channel Power)

    Measures the power and the power density within a specified channel bandwidth. The span and bandwidth are automatically set to smaller values in this measurement type.
    4. OBW (Occupied Bandwidth)
    Calculates the power within whole bandwidth by integral operation and works out the occupied bandwidth by this value based on the specified power ratio. The centre frequency difference between the measured channel and the analyser will also be given in the measurement results.
    5. EBW (Emission Bandwidth)
    Measures the bandwidth of the two points at both sides of the max signal when the amplitude of this max signal falls off X dB within the span range.
    6. C/N Ratio
    Measures the power of both the carrier and the noise within specified bandwidth and calculates their ratio.
    7. Harmo Dist (Harmonic Distortion)
    Measures each order harmonic power and THD (Total Harmonic Distortion) of carrier. The available range is up to 10 orders. And the fundamental wave amplitude of the carrier must be higher than -50 dBm or else the measurement will be invalid.
    8. TOI (Third Order Intermodulation)
    Measures the parameters of the TOI production generated by two signals which have the same amplitude and similar frequency. The measured results include the frequency and amplitude of the Base Lower, Base Upper, 3rd Order Lower and 3rd Order Upper signals, as well as the Intercept of both the 3rd Order Lower and 3rd Order Upper signals.
    9. Pass/Fail
    Compares the measured curve with the pre-edited, if the related rules are met, the result will be “Pass”, or else is “Fail”.
  • Is there an attenuator built-in on RIGOL’s DSA815?

    Is there an attenuator built-in on RIGOL’s DSA815?

    Yes, but you should still take care never to exceed the DSA815’s max input power or voltage (+20dBm or 100mW and 50VDC abs. max – pulsed waveforms can easily exceed this e.g. direct from a LISN – see below). The DSA internal attenuator function is variable to give relative 0-30dB and whilst it does not increase the permissible power, it is designed to allow the instrument to optimise its dynamic range, enabling it to measure louder signals without distortion and maintaining good accuracy.

    Note: A LISN isolates the power mains from the equipment under test, isolates any noise generated by the EUT, and couples the signals generated by the EUT to the spectrum analyser.

    Figure 3: Standard Conducted Emissions pre-compliance setup via a transient limiter. IMPORTANT: it is up to you to choose a limiter suited to your own system and testing regime and do your own checks first: transients can easily exceed the input rating of your analyser (e.g. at switch-on – it is generally regarded as good LISN practice to avoid switching on the LISN with any analyser connected!) A modern DSA’s internal level indicator is not intended to warn against fast transients and internal attenuators do not provide greater input circuit protection.
    CONCLUSION: if in doubt do your own measurements in any test environment on a resilient instrument (such as a suitably voltage-rated fast scope) first to ensure you always operate your more sensitive instruments, such as spectrum analysers, within the stated input limit envelope.

  • I just received mine… how can I tell if it’s working OK?

    I just received mine… how can I tell if it’s working OK?

    First follow the procedure for Factory Reset. Then check the noise performance is as it should be. You will need a carefully sequence of settings for this… use the quick video tip on our RIGOL pages.